OLYMPUS A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05,06 report with the FDA on 2008-01-30 for OLYMPUS A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[798594] The user facility returned the device to olympus for service. The device was received with a user's report of "no image and it burned a pt. " the user facility was contacted several times to obtain additional info regarding this report; however, no further info was obtained.
Patient Sequence No: 1, Text Type: D, B5


[7934524] The device was returned to olympus national service center for service. The device was then forwarded to ra for investigation due to the nature of the user's report. The investigation confirmed the user's report of "no image. " the outer tube of the telescope was severely dented, bent and bore multiple scratches. The impacts to the outer tube appeared to have caused image problems and were determined to be due to physical damage. The eval did not duplicate the device becoming unusually warm. Olympus will continue to work with the user facility to obtain clarifying info regarding the pt's outcome. If additional and significant info becomes available at a later date, a supplemental report will follow. This report is being submitted as an mdr in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2008-00009
MDR Report Key988725
Report Source00,05,06
Date Received2008-01-30
Date of Report2007-12-27
Date Added to Maude2008-07-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2008-01-30
Model NumberA50021A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key989666
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressHAMBURG


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.