IPC? HANDPIECE - INDIGO? DRILL 1845000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-27 for IPC? HANDPIECE - INDIGO? DRILL 1845000 manufactured by Medtronic Xomed Inc..

Event Text Entries

[185225886] The device was received for evaluation and repair. Analysis temperature testing found the device heated from 74? F to 158? F in the first few minutes. The temperature after 1 minute at 60k rpm was 158? F for the front, 126? F for the rear, and 73? F for t ambient. The bearings and motor were worn. The motor and motor collet were replaced. The device was repaired to manufacturing specifications. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185225887] Information was received from a healthcare professional (hcp) regarding a handpiece. It was reported that the handpiece was overheating. There was no patient impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00188
MDR Report Key9887331
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-01-26
Date Mfgr Received2020-01-27
Device Manufacturer Date2015-05-16
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIPC? HANDPIECE - INDIGO? DRILL
Generic NameDRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Product CodeHBE
Date Received2020-03-27
Returned To Mfg2020-02-11
Model Number1845000
Catalog Number1845000
Lot Number209592797
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report