LEVEL 1? HOTLINE? LOW FLOW SYSTEM HL-390 CON-HL-390

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-27 for LEVEL 1? HOTLINE? LOW FLOW SYSTEM HL-390 CON-HL-390 manufactured by Smiths Medical Asd; Inc..

Event Text Entries

[185193134] One hotline? Low flow system was returned for analysis. Visual inspection revealed that the front cover was cracked; confirming complaint. Based on the evidence, the root cause was found to be from impact to the corner of the cover.
Patient Sequence No: 1, Text Type: N, H10


[185193135] Information was received indicating that the front cover to a smiths medical level 1? Hotline? Low flow system was reported to be cracked. There were no reported adverse effects.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02243
MDR Report Key9887338
Report SourceUSER FACILITY
Date Received2020-03-27
Date of Report2020-03-26
Date Mfgr Received2020-02-26
Device Manufacturer Date2018-08-16
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS,, MN
Manufacturer CountryUS
Manufacturer Phone3833310
Manufacturer G1SMITHS MEDICAL ASD; INC.
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS,, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1? HOTLINE? LOW FLOW SYSTEM
Generic NameWARMER, THERMAL, INFUSION FLUID
Product CodeLGZ
Date Received2020-03-27
Returned To Mfg2019-08-27
Model NumberHL-390
Catalog NumberCON-HL-390
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD; INC.
Manufacturer Address6000 NATHAN LANE N MINNEAPOLIS,, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27