MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE manufactured by Alcon Research, Llc - Irvine Technology Center.
[186755648]
Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[186755649]
A physician reported three patient's experienced a rhexis break capsular rupture following a software update on the system. The surgeon noted that the ultrasounds were too "effective". This file represents the second of four patients. Additional information has been requested but not received. A completed questionnaire form the surgeon has been received. The cataract was a 4+ soft cortico-nuclear. The surgeon noted an occlusion during the procedure. Aspiration was too strong at the start of phacoemulsification chop. There was a rupture of anterior capsule. (rhexis) at 5 o? Clock position as soon as introduction of the handpiece. There was a problem positioning the toric intraocular lens (iol) because of the placement axis. The iol was placed in the capsule bag. The patient's symptoms have resolved.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2028159-2020-00283 |
MDR Report Key | 9888014 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date of Event | 2020-01-13 |
Date Mfgr Received | 2020-02-27 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. JONATHAN SCHLECH |
Manufacturer Street | 6201 SOUTH FREEWAY MAIL STOP AB2-6 |
Manufacturer City | FORT WORTH TX 76134 |
Manufacturer Country | US |
Manufacturer Postal | 76134 |
Manufacturer Phone | 8175514979 |
Manufacturer G1 | ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER |
Manufacturer Street | 15800 ALTON PARKWAY |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal Code | 92618 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE |
Generic Name | UNIT, PHACOFRAGMENTATION |
Product Code | HQC |
Date Received | 2020-03-27 |
Model Number | NA |
Catalog Number | ASKU |
Lot Number | ASKU |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER |
Manufacturer Address | 15800 ALTON PARKWAY IRVINE CA 92618 US 92618 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2020-03-27 |