MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-27 for SYNCHROMED II 8637-40 manufactured by Medtronic Puerto Rico Operations Co..
[186524945]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186524946]
Information was received from a healthcare professional (hcp) via a clinical study regarding a patient receiving gablofen 2000 mcg for a total dose of 805 mcg/day via an implantable pump. It was reported they were not able to refill or interrogate the pump. It was suspected and confirmed that the pump had flipped. There was no known factors that may have led or contributed to the issue. Diagnostics/troubleshooting included having an x-ray. Acti ons/interventions included surgical intervention where a pocket revision occurred for the flipped pump and the pump was placed back in. It was noted that the issue was resolved at time of report. The patient's status at time of report was alive-no injury. The patient's medical history was asked, but unknown. The event date was asked, but unknown. No further complications were reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004209178-2020-06250 |
MDR Report Key | 9888028 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date Mfgr Received | 2020-03-26 |
Device Manufacturer Date | 2017-02-21 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LISA WOODWARD CLARK |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635263920 |
Manufacturer G1 | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Street | ROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK |
Manufacturer City | JUNCOS PR 00777 |
Manufacturer Country | US |
Manufacturer Postal Code | 00777 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SYNCHROMED II |
Generic Name | PUMP, INFUSION, IMPLANTED, PROGRAMMABLE |
Product Code | LKK |
Date Received | 2020-03-27 |
Model Number | 8637-40 |
Catalog Number | 8637-40 |
Device Expiration Date | 2018-08-14 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Address | ROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-27 |