O-ARM 1000 IMAGING SYSTEM BI70000027120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-27 for O-ARM 1000 IMAGING SYSTEM BI70000027120 manufactured by Medtronic Navigation, Inc (littleton).

Event Text Entries

[185223925] Other relevant device(s) are: product id: kit svc bi71000514 pcba mvs pwr bd, serial/lot : unk. Product id: kit svc bi71000480 ups pwr sply, serial/lot : unk. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185223926] Medtronic received information regarding an imaging system being used prior to a procedure. It was reported that the site powered on the system, and them unplugged the mobile viewing station(mvs) from the wall to move it into the or. A minute after being unplugged, the mvs powered off and would not power back on, even after being plugged into a wall. After multiple attempts the site got the mvs to power on and when they went to connect the umbilical cord between the image acquisition system and the mvs, the mvs powered off and became unresponsive. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004785967-2020-00418
MDR Report Key9888544
Report SourceUSER FACILITY
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-25
Date Mfgr Received2020-03-25
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Street300 FOSTER ST
Manufacturer CityLITTLETON MA 01460
Manufacturer CountryUS
Manufacturer Postal Code01460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO-ARM 1000 IMAGING SYSTEM
Generic NameIMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Product CodeOXO
Date Received2020-03-27
Model NumberBI70000027120
Catalog NumberBI70000027120
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Address300 FOSTER ST LITTLETON MA 01460 US 01460


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27