LEVEL 1 HOTLINE LOW FLOW SYSTEM L-70NI

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,user facility report with the FDA on 2020-03-27 for LEVEL 1 HOTLINE LOW FLOW SYSTEM L-70NI manufactured by Smiths Medical Asd, Inc.

Event Text Entries

[185237284] Evaluation results: one level 1 hotline disposable was received for investigation in used condition. The sample was visually inspected at a distance of 12 to 24 inches and under normal conditions of illumination. The investigator noted that there was a deformation in the connector. The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution. No issues were identified that would have impacted this event. The investigator suspected that the connector reflux became damaged after the product left the manufacturing facility. A definitive root cause was not established however.
Patient Sequence No: 1, Text Type: N, H10


[185237285] It was reported that the twin-tube connectors were bent and sticking together. The operator was unable to insert the tube into the socket of the level 1 hotline low flow system (hl-90). The product problem was observed upon opening. No patient injury or complications were reported in relation to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02279
MDR Report Key9888910
Report SourceFOREIGN,OTHER,USER FACILITY
Date Received2020-03-27
Date of Report2020-03-27
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1 HOTLINE LOW FLOW SYSTEM
Generic NameWARMER, THERMAL, INFUSION FLUID
Product CodeLGZ
Date Received2020-03-27
Returned To Mfg2019-12-26
Model NumberL-70NI
Catalog NumberL-70NI
Lot Number3806399
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report