11CM ANGLE ATTACHMENT QD11

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for 11CM ANGLE ATTACHMENT QD11 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[185253125] Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. The actual device was returned for evaluation. During repair, an evaluation was performed and it was determined that the reported condition was confirmed. The assignable root cause was determined to be due to component failure from normal wear. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[185253126] It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device had heat and the bearings, tensioners and internal parts were damaged. It was further determined that the device failed pretest for vibration and temperature assessment. It was noted in the service order that the device bearings were worn. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown. However, it was reported that the event occurred in 2019. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045834-2020-00560
MDR Report Key9888995
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-02-03
Date of Event2019-01-01
Date Mfgr Received2020-03-18
Device Manufacturer Date2017-06-14
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KARA DITTY-BOVARD
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal33410
Manufacturer Phone6103142063
Manufacturer G1DEPUY SYNTHES PRODUCTS LLC
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal Code33410
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name11CM ANGLE ATTACHMENT
Generic NameMOTOR, DRILL, ELECTRIC - ATTACHMENT
Product CodeHBC
Date Received2020-03-27
Returned To Mfg2020-02-03
Model NumberQD11
Catalog NumberQD11
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer Address4500 RIVERSIDE DRIVE PALM BEACH GARDENS FL 33410 US 33410


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report