[186298764]
The reason for this revision surgery was reported as dislocation. The previous surgery and the surgery detailed in this event occurred 38 days apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at hospital and not made available to djo surgical for examination. A review of the device history records (dhr) shows that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint. The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to dislocation. There were no findings during this evaluation that indicate the reported devices were defective. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event. Due to short time between previous and revision surgery, it seems that the event may possibly occurred due to patient noncompliance with medical instruction, lack of care, improper implant selection, patient activities or trauma. There are no indications of a product or process issue affecting implant safety or effectiveness.
Patient Sequence No: 1, Text Type: N, H10
[186298765]
Revision surgery - baseplate disassociated with the glenoid causing the patient to dislocate.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1644408-2020-00261 |
MDR Report Key | 9889233 |
Report Source | COMPANY REPRESENTATIVE |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date of Event | 2020-02-28 |
Date Mfgr Received | 2020-02-28 |
Device Manufacturer Date | 2019-11-26 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | TEFFANY HUTTO |
Manufacturer Street | 9800 METRIC BLVD |
Manufacturer City | AUSTIN, TX 78758-5445, |
Manufacturer Country | US |
Manufacturer G1 | ENCORE MEDICAL L.P. |
Manufacturer Street | 9800 METRIC BLVD |
Manufacturer City | AUSTIN, TX 78758-5445, |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | RSP |
Generic Name | RSP BASEPLATE, 30MM, W/P2 COATING |
Product Code | PHX |
Date Received | 2020-03-27 |
Model Number | 508-32-204 |
Catalog Number | 508-32-204 |
Lot Number | 769P1730 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ENCORE MEDICAL L.P. |
Manufacturer Address | 9800 METRIC BLVD AUSTIN, TX 78758-5445, US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-27 |