G7 STR MONOBLOCK SHELL INSRTR N/A 110003450

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2020-03-27 for G7 STR MONOBLOCK SHELL INSRTR N/A 110003450 manufactured by Zimmer Biomet, Inc..

Event Text Entries

[186298392] (b)(4). Report source: (b)(6). Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted. Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-01353.
Patient Sequence No: 1, Text Type: N, H10


[186298393] It was reported that the inserter would not disengage from the shell. The shell was removed to try and remove the handle from the shell but they were unable to remove it. A second set was required to get another handle and a larger implant was required due to removing the original shell and making the acetabular hole larger when removing. Attempts have been made and additional information on the reported event is unavailable at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001825034-2020-01355
MDR Report Key9889350
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-18
Date Mfgr Received2020-03-18
Device Manufacturer Date2016-01-06
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER BIOMET, INC.
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameG7 STR MONOBLOCK SHELL INSRTR
Generic NameINSTRUMENT, HIP
Product CodePBI
Date Received2020-03-27
Model NumberN/A
Catalog Number110003450
Lot Number291120
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET, INC.
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.