CARDIOSAVE HYBRID INT TYPE E/F PLUG N/A 0998-00-0800-55

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for CARDIOSAVE HYBRID INT TYPE E/F PLUG N/A 0998-00-0800-55 manufactured by Datascope Corp. - Mahwah.

Event Text Entries

[185538468] The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date. A getinge field service engineer (fse) was dispatched to evaluate the iabp. The fse checked the iabp according to the leak troubleshooting procedure and found that the leak is located on the quick disconnect point. The fse noted that a new o-ring was installed and greased; however the issue persisted. The stm ordered the male and female connector to replace. Iabp was not cleared for clinical use. A supplemental report will be submitted when additional information is provided. Full initial reporter name: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[185538469] It was reported that during a routine check of a cardiosave intra-aortic balloon pump (iabp) a helium leak on the bottle side (the tank empties too fast). There was no patient involvement, thus no adverse event was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2249723-2020-00511
MDR Report Key9889410
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-03-25
Date Mfgr Received2020-03-05
Device Manufacturer Date2012-08-01
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street1300 MACARTHUR BLVD
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer G1DATASCOPE CORP. - MAHWAH
Manufacturer Street1300 MACARTHUR BLVD
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal Code07430
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARDIOSAVE HYBRID INT TYPE E/F PLUG
Generic NameSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Product CodeDSP
Date Received2020-03-27
Model NumberN/A
Catalog Number0998-00-0800-55
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDATASCOPE CORP. - MAHWAH
Manufacturer Address1300 MACARTHUR BLVD MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27