WATCHMAN TRUSEAL ACCESS SYSTEM M635TU70040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-27 for WATCHMAN TRUSEAL ACCESS SYSTEM M635TU70040 manufactured by Boston Scientific Corporation.

Event Text Entries

[185662953] It was reported the access system was damaged/defective. A left atrial appendage (laa) closure procedure was being performed. A watchman truseal access system (was) was positioned and a 27mm watchman laa closure device & delivery system (wds) were used. After entering the patient's body, the was was found to be kinked and also damaged the 27mm wds. Both the wds and was were removed and a new was and wds were prepped. After prepping the second was and wds, inserting the system as a whole into the patients body, and successfully deploying the second 27 mm device, it was noted that it seemed that something was hanging off the sheath in the left atrium. More heparin was given as thrombus formation was suspected; however, the heparin did not mitigate the issue. The 27mm device was released after going through pass criteria per dfu and the whole system was removed. Upon removal of the sheath, it was noted that a fiber-like strand was hanging off the end of the sheath. No harm was done to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03795
MDR Report Key9889436
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-06-28
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATCHMAN TRUSEAL ACCESS SYSTEM
Generic NameSYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Product CodeDQY
Date Received2020-03-27
Model NumberM635TU70040
Catalog NumberM635TU70040
Lot Number0024033040
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report