ACTIV.A.C.? THERAPY SYSTEM WNDACT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-27 for ACTIV.A.C.? THERAPY SYSTEM WNDACT manufactured by Kinetic Concepts, Inc..

Event Text Entries

[185272767] Based on the information obtained regarding the device, kci found evidence that the device had a collapsed dome. There is no injury associated with this event. Kci is reporting this event found during servicing of the unit as a device malfunction that has the potential to result in injury if it were to recur.
Patient Sequence No: 1, Text Type: N, H10


[185272768] On (b)(6) 2020, kci technical services identified a power switch with collapsed dome. On (b)(6) 2020, kci quality engineering performed an evaluation of the device. Inspection of the device found the power button was collapsed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009897021-2020-00134
MDR Report Key9889501
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-02-26
Date Mfgr Received2020-02-26
Device Manufacturer Date2011-02-03
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEVEN JACKSON
Manufacturer Street6203 FARINON DRIVE
Manufacturer CitySAN ANTONIO, TX
Manufacturer CountryUS
Manufacturer Phone2556438
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACTIV.A.C.? THERAPY SYSTEM
Generic NameOMP
Product CodeOMP
Date Received2020-03-27
Returned To Mfg2020-02-19
Model NumberWNDACT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKINETIC CONCEPTS, INC.
Manufacturer AddressSAN ANTONIO, TX US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.