PULSE GEN MODEL 106

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for PULSE GEN MODEL 106 manufactured by Cyberonics - Houston.

Event Text Entries

[188030131] During a generator battery replacement for normal expected battery depletion, the replacement m106 generator resulted in high impedance once connected to the lead. It was noted that the generator header hole was too small. The generator was sent back for product analysis, but the device has not been received by the manufacturer to date. No other relevant information has been received to date.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644487-2020-00514
MDR Report Key9889558
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-02
Date Mfgr Received2020-03-02
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RACHEL KOHN
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal77058
Manufacturer Phone2812287200
Manufacturer G1CYBERONICS - HOUSTON
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal Code77058
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePULSE GEN MODEL 106
Generic NameGENERATOR
Product CodeLYJ
Date Received2020-03-27
Model Number106
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCYBERONICS - HOUSTON
Manufacturer Address100 CYBERONICS BLVD HOUSTON TX 77058 US 77058


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27