BATTERY REAMER/DRILL FOR BPL II 530.705

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for BATTERY REAMER/DRILL FOR BPL II 530.705 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[186530976] Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. During repair, it was determined that the reported condition was confirmed. The assignable root cause was determined to be due to maintenance deficiency. Udi: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186530977] It was reported from (b)(6) that during service and evaluation, it was determined that the battery reamer device moving parts did not move smoothly - trigger. It was further determined that the device failed pretest for trigger. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown. However, it was reported that the event occurred in 2020. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030965-2020-02330
MDR Report Key9889730
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-03-18
Date of Event2020-01-01
Date Mfgr Received2020-03-18
Device Manufacturer Date2013-10-16
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KARA DITTY-BOVARD
Manufacturer StreetEIMATTSTRASSE 3
Manufacturer CityOBERDORF 4436
Manufacturer CountrySZ
Manufacturer Postal4436
Manufacturer Phone6103142063
Manufacturer G1SYNTHES PRODUKTIONS GMBH
Manufacturer StreetHAUPTSTRASSE 24
Manufacturer CityWALDENBURG 4437
Manufacturer CountrySZ
Manufacturer Postal Code4437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBATTERY REAMER/DRILL FOR BPL II
Generic NameMOTOR, SURGICAL INSTRUMENT, AC-POWERED
Product CodeGEY
Date Received2020-03-27
Returned To Mfg2020-03-09
Catalog Number530.705
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer AddressEIMATTSTRASSE 3 OBERDORF 4436 SZ 4436


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.