HOTLINE 3 BLOOD AND FLUID WARMER CON-HL-390

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-27 for HOTLINE 3 BLOOD AND FLUID WARMER CON-HL-390 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[185268265] Returned device was received in decent physical condition. The enclosure, tank cover, front cover, and faded line cord were all damaged. During the evaluation of the device, the tank was filled with water, the temp check was attached and the in line cord was plugged in and the power switch was turned on. It was found that water was leaking from the lower right corner of the tank cover. The tank cover was found to be the cause of the complaint. This will be replaced. The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution. No issues were identified that would have impacted this event.
Patient Sequence No: 1, Text Type: N, H10


[185268266] Information was received indicating that a smiths medical fluid warmer had a leak. No adverse events were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02372
MDR Report Key9889782
Report SourceUSER FACILITY
Date Received2020-03-27
Date of Report2020-03-27
Date Mfgr Received2020-02-27
Device Manufacturer Date2012-03-22
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer Phone3833310
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHOTLINE 3 BLOOD AND FLUID WARMER
Generic NameWARMER, THERMAL, INFUSION FLUID
Product CodeLGZ
Date Received2020-03-27
Returned To Mfg2019-09-11
Catalog NumberCON-HL-390
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.