MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-27 for CS300 N/A 0998-00-3023-53 manufactured by Datascope Corp. - Mahwah.
[185543092]
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date. (b)(6). A getinge service territory manager (stm) was dispatched to evaluate the iabp unit. The stm checked the fiber-optic module multiple times with a tester and could not reproduce the issue. The safety disk was replaced and the compressor was rebuilt as part of a preventative maintenance (pm). The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer.
Patient Sequence No: 1, Text Type: N, H10
[185543093]
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) would not calibrate. There was no alarm or log entry for this incident. There was no harm or injury to the patient and no adverse event was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2249723-2020-00512 |
MDR Report Key | 9889904 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-27 |
Date of Report | 2020-03-24 |
Date of Event | 2020-03-02 |
Date Mfgr Received | 2020-03-06 |
Device Manufacturer Date | 2007-05-01 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 1300 MACARTHUR BLVD |
Manufacturer City | MAHWAH NJ 07430 |
Manufacturer Country | US |
Manufacturer Postal | 07430 |
Manufacturer G1 | DATASCOPE CORP. - MAHWAH |
Manufacturer Street | 1300 MACARTHUR BLVD |
Manufacturer City | MAHWAH NJ 07430 |
Manufacturer Country | US |
Manufacturer Postal Code | 07430 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CS300 |
Generic Name | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL |
Product Code | DSP |
Date Received | 2020-03-27 |
Model Number | N/A |
Catalog Number | 0998-00-3023-53 |
Lot Number | N/A |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DATASCOPE CORP. - MAHWAH |
Manufacturer Address | 1300 MACARTHUR BLVD MAHWAH NJ 07430 US 07430 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-27 |