MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for MINICAP TRANSFER SET 5C4482 manufactured by Baxter Healthcare Corporation.
[185295958]
(b)(6). The device was received for evaluation. A visual inspection was performed, and it was noted that the returned transfer set had a broken occlude foot which would had caused the reported condition of fluid flow while the twist clamp was closed. The reported condition was verified. The cause of the broken occlude foot could not be determined. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
[185295959]
It was reported that a peritoneal dialysis (pd) transfer set was leaking; further described as? Fluid flow with the twist clamp closed?. The customer observed fluid flow while the twist clamp on the transfer set was closed during pd therapy. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1416980-2020-01794 |
MDR Report Key | 9890593 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date of Event | 2020-03-03 |
Date Mfgr Received | 2020-03-03 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 25212 W. ILLINOIS ROUTE 120 |
Manufacturer City | ROUND LAKE IL 60073 |
Manufacturer Country | US |
Manufacturer Postal | 60073 |
Manufacturer Phone | 2242702068 |
Manufacturer G1 | BAXTER HEALTHCARE - MOUNTAIN HOME |
Manufacturer Street | 1900 N HIGHWAY 201 |
Manufacturer City | MOUNTAIN HOME AR 72653 |
Manufacturer Country | US |
Manufacturer Postal Code | 72653 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MINICAP TRANSFER SET |
Generic Name | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE |
Product Code | KDJ |
Date Received | 2020-03-27 |
Returned To Mfg | 2020-03-16 |
Model Number | NA |
Catalog Number | 5C4482 |
Lot Number | ASKU |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BAXTER HEALTHCARE CORPORATION |
Manufacturer Address | DEERFIELD IL |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-27 |