MINICAP TRANSFER SET 5C4482

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for MINICAP TRANSFER SET 5C4482 manufactured by Baxter Healthcare Corporation.

Event Text Entries

[185295958] (b)(6). The device was received for evaluation. A visual inspection was performed, and it was noted that the returned transfer set had a broken occlude foot which would had caused the reported condition of fluid flow while the twist clamp was closed. The reported condition was verified. The cause of the broken occlude foot could not be determined. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[185295959] It was reported that a peritoneal dialysis (pd) transfer set was leaking; further described as? Fluid flow with the twist clamp closed?. The customer observed fluid flow while the twist clamp on the transfer set was closed during pd therapy. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2020-01794
MDR Report Key9890593
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Street1900 N HIGHWAY 201
Manufacturer CityMOUNTAIN HOME AR 72653
Manufacturer CountryUS
Manufacturer Postal Code72653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMINICAP TRANSFER SET
Generic NameSET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Product CodeKDJ
Date Received2020-03-27
Returned To Mfg2020-03-16
Model NumberNA
Catalog Number5C4482
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.