ARCHITECT I2000SR ANALYZER 3M74-02 03M74-02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-27 for ARCHITECT I2000SR ANALYZER 3M74-02 03M74-02 manufactured by Abbott Laboratories (irving Ia/cc).

Event Text Entries

[188259143] During the subsequent site visit by the field service engineer (fse) the analyzer was inspected, and a part replacement was performed. A review of the analyzer service history identified no contributing factors to the current complaint. There have been no further occurrences of discrepant results after the part was replaced by the fse. A review of the immunoassay systems tracking and trending data revealed no systemic issues or trends associated with the discrepant result described in this complaint. Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue. A review of the product labeling concluded that the issue is sufficiently addressed. Based on the investigation no product deficiency was identified for the architect i2000sr serial number (b)(4). All available patient information was included. Additional patient details are not available.
Patient Sequence No: 1, Text Type: N, H10


[188259144] The customer reported a falsely elevated architect stat troponin-i result on a (b)(6)-yr old female patient. Results provided: (b)(6) 2020 sid 007309341 = 0. 35 / 0. 00 ng/ml, another instrument = 0. 07 ng/ml. Customer's negative range: 0. 03-0. 19 ng/ml. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3016438761-2020-00015
MDR Report Key9890759
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-14
Date Mfgr Received2020-03-16
Device Manufacturer Date2014-06-01
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTIAN LEE
Manufacturer Street100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224668-294
Manufacturer G1ABBOTT LABORATORIES (IRVING IA/CC)
Manufacturer Street1915 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT I2000SR ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeJJE
Date Received2020-03-27
Model Number3M74-02
Catalog Number03M74-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES (IRVING IA/CC)
Manufacturer Address1915 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2024 FDA.report