MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-27 for GRAFTMASTER 1012580-19 manufactured by Abbott Vascular.
[188552686]
The device was not returned for analysis. A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint. Additionally, a review of the complaint history did not indicate a lot specific quality issue. The investigation determined the reported failure to advance appears to be related to circumstances of the procedure. There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device. The other graftmaster is filed under a separate medwatch report number.
Patient Sequence No: 1, Text Type: N, H10
[188552687]
It was reported that the procedure was to treat a perforation caused by another device. The first graftmaster failed to seal as it was unable to reach the entire area of perforation due to proximal vessel tortuosity. Another same size graftmaster was advanced but failed to cross due to the tortuosity of the vessel. The perforation was sealed using a bare metal stent. There was no reported adverse patient sequela or a clinically significant delay in procedure. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2024168-2020-03042 |
MDR Report Key | 9890790 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date of Event | 2020-02-19 |
Date Mfgr Received | 2020-03-05 |
Device Manufacturer Date | 2018-11-01 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. LINDSEY BELL |
Manufacturer Street | 26531 YNEZ RD. |
Manufacturer City | TEMECULA CA 925914628 |
Manufacturer Country | US |
Manufacturer Postal | 925914628 |
Manufacturer Phone | 9519143996 |
Manufacturer G1 | ABBOTT VASCULAR, REG # 3005718570 (P099) |
Manufacturer Street | CASHEL ROAD |
Manufacturer City | CLONMEL TIPPERARY |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GRAFTMASTER |
Generic Name | CORONARY STENT DELIVERY SYSTEM |
Product Code | MAF |
Date Received | 2020-03-27 |
Model Number | 1012580-19 |
Catalog Number | 1012580-19 |
Lot Number | 8110141 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT VASCULAR |
Manufacturer Address | 26531 YNEZ RD. TEMECULA CA 925914628 US 925914628 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-27 |