COLIBRI II 532.101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for COLIBRI II 532.101 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[186526021] Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. The actual device was returned for evaluation. The small battery drive device was evaluated and the reported condition that the device was not working was not confirmed. Therefore, an assignable root cause was not determined. However, during evaluation it was observed that the device had a sticky trigger and failed pre-test for check for sticky speed trigger. The assignable root cause of this condition was determined to be traced to component failure due to normal wear. Udi? (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186526022] It was reported from (b)(6) that during service and evaluation, it was determined that the small battery drive device had insufficient/low power. It was further determined that the device failed pretest for check function forward/reverse, check the power with test bench, and check for sticky speed trigger. It was noted in the service order that the device was not working. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was not reported, however, it was reported that the event occurred in 2020. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030965-2020-02309
MDR Report Key9890991
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-02-26
Date of Event2020-01-01
Date Mfgr Received2020-03-23
Device Manufacturer Date2014-07-28
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KARA DITTY-BOVARD
Manufacturer StreetEIMATTSTRASSE 3
Manufacturer CityOBERDORF 4436
Manufacturer CountrySZ
Manufacturer Postal4436
Manufacturer Phone6103142063
Manufacturer G1SYNTHES PRODUKTIONS GMBH
Manufacturer StreetHAUPTSTRASSE 24
Manufacturer CityWALDENBURG 4437
Manufacturer CountrySZ
Manufacturer Postal Code4437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOLIBRI II
Generic NameINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Product CodeHWE
Date Received2020-03-27
Returned To Mfg2020-03-03
Catalog Number532.101
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer AddressEIMATTSTRASSE 3 OBERDORF 4436 SZ 4436


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.