AMPLATZER CRIBRIFORM OCCLUDER 9-ASD-MF-025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-27 for AMPLATZER CRIBRIFORM OCCLUDER 9-ASD-MF-025 manufactured by Aga Medical Corporation.

Event Text Entries

[185324754] Further information regarding this event has been requested. The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
Patient Sequence No: 1, Text Type: N, H10


[185324755] On (b)(6) 2020, a 25mm amplatzer cribriform occluder (lot#7262218) was selected for implant. The device was prepared in the normal procedure. Once inside the patient, distortion was seen on the disc when opened on the left atrium side. The same deformity was seen on the disc when opened on the right atrium side. Another 25mm amplatzer cribriform occluder (lot#7253343) was selected for implant, but the same deformation was seen during preparation. The device did not come in contact with the patient. A 25mm amplatzer pfo occluder was then selected for implant and the process was completed successfully. The patient remained stable throughout the procedure and there was no clinically significant delay in the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2135147-2020-00098
MDR Report Key9891567
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2019-02-28
Date Mfgr Received2020-02-28
Device Manufacturer Date2019-11-07
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1AGA MEDICAL CORPORATION
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLATZER CRIBRIFORM OCCLUDER
Generic NameTRANSCATHETER SEPTAL OCCLUDER
Product CodeMLV
Date Received2020-03-27
Model Number9-ASD-MF-025
Catalog Number9-ASD-MF-025
Lot Number7262218
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAGA MEDICAL CORPORATION
Manufacturer Address5050 NATHAN LANE NORTH PLYMOUTH MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-27