MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-27 for SMALL PASSIVE CRANIAL REFERENCE FRAME 961-337 manufactured by Medtronic Navigation, Inc.
[186751422]
Analysis of the returned reference frame found no fault. There was no apparent physical damage and none of the posts were bent. The reference frame returned good geometry, divot error, and normal tracking. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186751443]
Medtronic received information regarding a navigation system that was used outside of procedure. The site had issues with tracking. The passive tracking spheres were changed without resolution. It was noted some of the posts on the instruments were bent, which threw off the geometry error. No patient present.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1723170-2020-01095 |
MDR Report Key | 9891859 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-27 |
Date of Report | 2020-03-27 |
Date of Event | 2020-03-09 |
Date Mfgr Received | 2020-03-09 |
Device Manufacturer Date | 2011-12-16 |
Date Added to Maude | 2020-03-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACY RUEMPING |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635260594 |
Manufacturer G1 | MEDTRONIC NAVIGATION, INC |
Manufacturer Street | 826 COAL CREEK CIRCLE |
Manufacturer City | LOUISVILLE CO 80027 |
Manufacturer Country | US |
Manufacturer Postal Code | 80027 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SMALL PASSIVE CRANIAL REFERENCE FRAME |
Generic Name | INSTRUMENT, STEREOTAXIC |
Product Code | HAW |
Date Received | 2020-03-27 |
Returned To Mfg | 2020-03-17 |
Model Number | 961-337 |
Catalog Number | 961-337 |
Lot Number | 111216 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC NAVIGATION, INC |
Manufacturer Address | 826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-27 |