MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2020-03-27 for RENASYS GO 66021496 manufactured by Smith & Nephew Medical Ltd..

Event Text Entries

[185901688] It was reported that the patient stated that his device has been displaying full blockage when there is no drainage in the canister at all. After the patient performed this first troubleshooting, the "canister full blockage alarm" did not get resolved. The patient stated that every time his device displays full blockage, he would still feel the suctioning motion and could feel that the device is still working. He confirmed that the tubing is never kinked or bent. After the patient performed this last troubleshooting step, the full blockage alarm got resolved. Informed the patient that since the alarm condition was resolved, the blockage was present within the soft port and per clinical guideline, this needs to be reassessed and replaced as needed. A back up device was not available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043484-2020-00225
MDR Report Key9892105
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-03-02
Date Mfgr Received2020-03-02
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR SARAH FREESTONE
Manufacturer Street101 HESSLE ROAD
Manufacturer CityHULL HU3 2BN
Manufacturer CountryUK
Manufacturer PostalHU3 2BN
Manufacturer Phone0447940038
Manufacturer G1SMITH & NEPHEW MEDICAL LTD.
Manufacturer Street101 HESSLE ROAD
Manufacturer CityHULL HU3 2BN
Manufacturer CountryUK
Manufacturer Postal CodeHU3 2BN
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRENASYS GO
Generic NameNEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Product CodeOMP
Date Received2020-03-27
Catalog Number66021496
OperatorLAY USER/PATIENT
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW MEDICAL LTD.
Manufacturer Address101 HESSLE ROAD HULL HU3 2BN UK HU3 2BN


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27