PORTEX BIVONA TRACHEOSTOMY TUBE 67PFSS40

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-03-27 for PORTEX BIVONA TRACHEOSTOMY TUBE 67PFSS40 manufactured by Smiths Medical Asd; Inc..

Event Text Entries

[185480576] One bivona tracheostomy (trach) tube was returned for analysis in used conditions. The tube was visually inspected; no discrepancies noted. The sample was filled with air; noting the cuff inflated uniformly. The cuff was inflated and deflated 4 additional times; the cuff inflated properly each time. The manufacturing process and assembly process were both reviewed; no discrepancies. Verification of manufacturing process of the air way and balloon was performed on 32 units; all performed correctly. Based on the evidence there was no fault found as the complaint was not confirmed.
Patient Sequence No: 1, Text Type: N, H10


[185480577] Information was received indicating that the cuff to a smiths medical portex bivona tracheostomy (trach) tube was reported to not stay inflated during use. The trach was returned for analysis. There were no reported adverse effects.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02375
MDR Report Key9892389
Report SourceDISTRIBUTOR
Date Received2020-03-27
Date of Report2020-03-27
Date of Event2020-02-21
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS,, MN
Manufacturer CountryUS
Manufacturer Phone3833310
Manufacturer G1SMITHS MEDICAL ASD; INC.
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS,, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePORTEX BIVONA TRACHEOSTOMY TUBE
Generic NameTUBE TRACHEOSTOMY AND TUBE CUFF
Product CodeJOH
Date Received2020-03-27
Returned To Mfg2020-03-04
Model Number67PFSS40
Catalog Number67PFSS40
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD; INC.
Manufacturer Address6000 NATHAN LANE N MINNEAPOLIS,, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.