KYPHON V OSTEO INTRODUCER 13GA TROCAR/BEVEL VPT13D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-28 for KYPHON V OSTEO INTRODUCER 13GA TROCAR/BEVEL VPT13D manufactured by Medtronic Mexico.

Event Text Entries

[189090395] Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[189090396] It was reported that intra-op, trocar got stuck in the vertebral body, possibly due to heavy malleting and denser than normal bone. Upon attempting to pull out trocar, the purple hub on superficial end ripped off from the trocar placed in the pedicle and vertebral body. At times, physician was pulling so hard on the needle that the patient was being lifted off of the table. Once the purple hub ripped off, the surgeon had to use vice grips and a mallet to remove the needle from patient to prevent an open spine surgery to remove. The doctor was sure that the bond between the hub and cannula became loose due to heavy malleting. The patient had hematoma at access site, possibly due to excessive pulling force on the trocar and pedicle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612164-2020-01385
MDR Report Key9892527
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-28
Date of Report2020-03-28
Date of Event2020-03-09
Date Mfgr Received2020-03-09
Device Manufacturer Date2018-11-30
Date Added to Maude2020-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACIE ZIEMBA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1MEDTRONIC MEXICO
Manufacturer StreetAV. PASEO DEL CUCAPAH #10510
Manufacturer CityTIJUANA,BC 22570
Manufacturer CountryMX
Manufacturer Postal Code22570
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKYPHON V OSTEO INTRODUCER 13GA TROCAR/BEVEL
Generic NameORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Product CodeLXH
Date Received2020-03-28
Model NumberNA
Catalog NumberVPT13D
Lot Number216656426
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC MEXICO
Manufacturer AddressAV. PASEO DEL CUCAPAH #10510 TIJUANA,BC 22570 MX 22570


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.