IPC? HANDPIECE - XPS? STRAIGHTSHOT? M4 1898200TRF

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-28 for IPC? HANDPIECE - XPS? STRAIGHTSHOT? M4 1898200TRF manufactured by Medtronic Xomed Inc..

Event Text Entries

[185494488] Analysis found that the customer complaint of overheating could not be confirmed since the device was not tested for pre-temperature test. However, it was found that the handpiece was stalling but was working. All the internal parts were corroded. There was foreign material built-up on rear bearings and the cable was cut. Removed the foreign material and replaced with new rear seals and bearing and motor cable assembly. Also replaced old serial number (b)(4) with new udi serial (b)(4). The device was repaired, cleaned, tested and passed all manufacturing specifications. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185494489] A health care provider (hcp) reported that the handpiece was not working properly. There was no patient impact. On follow-up, it was reported that the handpiece was running hot during set-up.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00189
MDR Report Key9893137
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-28
Date of Report2020-03-28
Date of Event2020-03-03
Date Mfgr Received2020-03-06
Device Manufacturer Date2013-07-23
Date Added to Maude2020-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIPC? HANDPIECE - XPS? STRAIGHTSHOT? M4
Generic NameDRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Product CodeERL
Date Received2020-03-28
Returned To Mfg2020-03-12
Model Number1898200TRF
Catalog Number1898200TRF
Lot Number207217376
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-28

© 2024 FDA.report