AMPLATZER SEPTAL OCCLUDER 9-ASD-016

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-29 for AMPLATZER SEPTAL OCCLUDER 9-ASD-016 manufactured by Aga Medical Corporation.

Event Text Entries

[185441060] An event of the occluder appearing "constrained" while deployed and the septum and deploying deformed outside the body was reported. The investigation confirmed the device met functional and dimensional specifications when analyzed at abbott. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications. The cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185441061] On (b)(6) 2020, a 16mm asd device was advanced in usual fashion to the inter atrial septum after balloon sizing revealed a 15mm asd. The device was deployed on the septum however the occluder appeared to be constrained while deployed correctly on the inter atrial septum. The device was recaptured and removed and exchanged for a 14mm asd device which was subsequently deployed successfully and released and remains implanted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2135147-2020-00127
MDR Report Key9895068
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-29
Date of Report2020-03-29
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-07-05
Date Added to Maude2020-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1AGA MEDICAL CORPORATION
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLATZER SEPTAL OCCLUDER
Generic NameTRANSCATHETER SEPTAL OCCLUDER
Product CodeMLV
Date Received2020-03-29
Returned To Mfg2020-03-12
Model Number9-ASD-016
Catalog Number9-ASD-016
Lot Number7076148
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAGA MEDICAL CORPORATION
Manufacturer Address5050 NATHAN LANE NORTH PLYMOUTH MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-29