466FXXXX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2020-03-30 for 466FXXXX manufactured by Cordis Corporation.

Event Text Entries

[188510435] The catalog number is unknown, if received it will be provided. Complaint conclusion: as reported, the patient underwent placement of an optease retrievable vena cava filter. The indication for the filter placement was not reported. At some point after the filter implantation, the patient became aware that the filter had tilted. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease retrievable vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Without images or procedural films for review, the reported filter tilt could not be confirmed and the exact cause could not be determined. Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness. Additionally, the timing and mechanism of the filter tilt is unknown. Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
Patient Sequence No: 1, Text Type: N, H10


[188510436] As reported by the legal brief, the patient underwent placement of a optease vena cava filter. The filter subsequently malfunctioned and caused injury and damages including, but not limited to: tilt.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1016427-2020-03968
MDR Report Key9896776
Report SourceOTHER
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-04
Date Mfgr Received2020-03-05
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARLA CASTRO
Manufacturer Street14201 NW 60TH AVE
Manufacturer CityMIAMI LAKES FL 33014
Manufacturer CountryUS
Manufacturer Postal33014
Manufacturer Phone7863138372
Manufacturer G1CORDIS CORPORATION
Manufacturer Street14201 NW 60TH AVE
Manufacturer CityMIAMI LAKES FL 33014
Manufacturer CountryUS
Manufacturer Postal Code33014
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name466FXXXX
Generic NameFILTER, INTRAVASCULAR, CARDIOVASCULAR
Product CodeDTK
Date Received2020-03-30
Model Number466FXXXX
Catalog Number466FXXXX
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCORDIS CORPORATION
Manufacturer Address14201 NW 60 AVENUE MIAMI LAKES FL 33014 US 33014


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2020-03-30

© 2024 FDA.report