STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 3 01.12.023

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-30 for STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 3 01.12.023 manufactured by Medacta International Sa.

Event Text Entries

[188223750] Batch review performed by medacta regulatory affairs department on 13 march 2020: lot 163265: (b)(4) items manufactured and released on 02-sep-2016. Expiration date: 2021-08-18. No anomalies found related to the problem. To date, (b)(4) items of the same lot have been already sold without any similar event reported. Clinical evaluation performed by medical affairs manager: partial hip revision surgery (stem and head) performed 3 years after cementless total hip arthroplasty in a (b)(6) year old farmer. Radiological images provided show the presence of radiolucent lines in gruen zones 1, 2, 6 and 7 and signs of stress shielding suggesting stem mobilization. Aseptic loosening is a possible literature described adverse event after cementless total hip arthroplasty and causes are often unknown. Early return to high level of activities may reduce primary stability of the stem and hinder implant fixation. No conclusion can be drawn on the basis of available information.
Patient Sequence No: 1, Text Type: N, H10


[188223751] Revision surgery performed 3 years and 1 month after the primary surgery due to the proximal loosening of a quadra h stem. The surgeon revised the stem and ball head. The patient was a heavy working farmer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005180920-2020-00203
MDR Report Key9896985
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-04
Date Mfgr Received2020-03-03
Device Manufacturer Date2016-09-02
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR STEFANO BAJ
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, SWITZERLAND 6874
Manufacturer CountrySZ
Manufacturer Postal6874
Manufacturer G1MEDACTA INTERNATIONAL SA
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, 6874
Manufacturer CountrySZ
Manufacturer Postal Code6874
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 3
Generic NameCEMENTLESS HIP STEM
Product CodeJDI
Date Received2020-03-30
Model Number01.12.023
Catalog Number01.12.023
Lot Number163265
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDACTA INTERNATIONAL SA
Manufacturer AddressSTRADA REGINA CASTEL SAN PIETRO, 6874 SZ 6874


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-30