O-ARM 1000 IMAGING SYSTEM 9732719

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-30 for O-ARM 1000 IMAGING SYSTEM 9732719 manufactured by Medtronic Navigation, Inc (littleton).

Event Text Entries

[185522048] No patient involved. Unique device identifier (udi) is unavailable. Other relevant device(s) are: product id: b i71000489, serial/lot #: unknown. A manufacturer representative went to the site to test the imaging system. It was noted system would shutdown quickly once driven. Remote desktop showed motion battery voltage dropping quickly and randomly. The system control board was replaced which resolved the issue. The system then passed checkout and performed as intended. Device manufacturing date is unavailable. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185522049] Medtronic received information regarding an imaging system used outside of procedure. It was reported that the system displayed a low battery message. The site will leave the system plugged into wall power to charge, once charged they will unplug to see if battery backup occurs. No patient present.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004785967-2020-00421
MDR Report Key9897067
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-20
Date Mfgr Received2020-03-20
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Street300 FOSTER ST
Manufacturer CityLITTLETON MA 01460
Manufacturer CountryUS
Manufacturer Postal Code01460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameO-ARM 1000 IMAGING SYSTEM
Generic NameIMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Product CodeOXO
Date Received2020-03-30
Model Number9732719
Catalog Number9732719
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC (LITTLETON)
Manufacturer Address300 FOSTER ST LITTLETON MA 01460 US 01460


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report