Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1IMPLANTABLE COLLAMER LENS (ICL)PHAKIC TORIC INTRAOCULAR LENSSTAAR SURGICAL COMPANYQCBVTICMO13.2NAN/AR N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12020-03-3001. R

Event Narratives#

N

Patient 1

THIS PRODUCT IS NOT MARKETED IN THE US. (B)(4).

D

Patient 1

THE REPORTER INDICATED THAT A 13.2MM, VTICMO13.2, -10.0/1.0/087 (SPHERE/CYLINDER/AXIS) IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO PATIENTS RIGHT EYE (OD) ON (B)(6) 2019. ON (B)(6) 2019 THE LENS WAS EXCHANGED FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THIS EXCHANGE RESOLVED THE PROBLEM.