AMS INFLATABLE PENILE PROSTHESIS 72404310

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-30 for AMS INFLATABLE PENILE PROSTHESIS 72404310 manufactured by Boston Scientific Corporation.

Event Text Entries

[186591289] It was reported that the patient complained of the pump not working since (b)(6) 2020 due to fluid loss with an inflatable penile prosthesis (ipp). At the time of check-up on (b)(6) 2020, the pump still was not functioning and the position was too high. The patient underwent a revision procedure, the physician explanted the ipp pump and a new ipp pump was implanted. During the revision procedure the physician noted a microscopic hole in the tube side of the pump, the cylinder and reservoir were re-filled and the procedure was completed. The patient recovered following the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183959-2020-01688
MDR Report Key9897859
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-11
Date Mfgr Received2020-03-12
Device Manufacturer Date2019-07-24
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALYSON HARRIS
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone4089353452
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal Code55311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS INFLATABLE PENILE PROSTHESIS
Generic NameDEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Product CodeFHW
Date Received2020-03-30
Model Number72404310
Catalog Number72404310
Lot Number1000307766
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address10700 BREN ROAD W MINNETONKA MN 55343 US 55343


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-30