MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-30 for O-ARM BI70000028120 manufactured by Medtronic Navigation, Inc (littleton).
[185524651]
A medtronic representative went to the site to test the equipment. Testing revealed the battery appeared to have overheated due to corrosion of battery terminal causing resistance/poor contact. All 38 system batteries, tray 4, contact terminals and wiring were replaced. The system then passed the system checkout and was found to be fully functional. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[185524652]
Medtronic received information regarding an imaging system being used outside of a procedure. It was reported that the image acquisition system (ias) had a thermal event. The biomedical engineer pulled out the middle battery on the top tray which was overheating to resolved the issue. This took place in a storage room. There was no patient involved.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004785967-2020-00424 |
MDR Report Key | 9897990 |
Report Source | COMPANY REPRESENTATIVE,USER F |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-03-09 |
Date Mfgr Received | 2020-03-09 |
Device Manufacturer Date | 2014-11-22 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACY RUEMPING |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635260594 |
Manufacturer G1 | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Street | 300 FOSTER ST |
Manufacturer City | LITTLETON MA 01460 |
Manufacturer Country | US |
Manufacturer Postal Code | 01460 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | O-ARM |
Generic Name | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE |
Product Code | OXO |
Date Received | 2020-03-30 |
Model Number | BI70000028120 |
Catalog Number | BI70000028120 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Address | 300 FOSTER ST LITTLETON MA 01460 US 01460 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |