MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-30 for FAN SPRAY KIT N/A 00515047501 manufactured by Zimmer Surgical, Inc..
[186723026]
This event has been recorded by zimmer biomet under (b)(4). Udi: (b)(4). Evaluation and investigation is in process. Once the investigation is complete, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10
[186723027]
It was reported that during the surgery, this complaint product did not work at all on high mode from the start of use. Upon inspection of the device it was discovered that the foreign substance which looks like the deposit from the electrolyte has be attached on the electrodes of the 1 of 8 batteries. No adverse events were reported as a result of this malfunction.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0001526350-2020-00324 |
MDR Report Key | 9897998 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-03-04 |
Date Mfgr Received | 2020-03-05 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. CHRISTINA ARNT |
Manufacturer Street | 56 E. BELL DRIVE |
Manufacturer City | WARSAW IN 46582 |
Manufacturer Country | US |
Manufacturer Postal | 46582 |
Manufacturer Phone | 5745273773 |
Manufacturer G1 | ZIMMER SURGICAL, INC. |
Manufacturer Street | 200 WEST OHIO AVENUE |
Manufacturer City | DOVER OH 44622 |
Manufacturer Country | US |
Manufacturer Postal Code | 44622 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | N/A |
Event Type | 3 |
Type of Report | 3 |
Brand Name | FAN SPRAY KIT |
Generic Name | LAVAGE, JET |
Product Code | FQH |
Date Received | 2020-03-30 |
Model Number | N/A |
Catalog Number | 00515047501 |
Lot Number | 64245126 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ZIMMER SURGICAL, INC. |
Manufacturer Address | 200 WEST OHIO AVENUE DOVER OH 44622 US 44622 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |