HT COMMAND 18 1013731

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-30 for HT COMMAND 18 1013731 manufactured by Abbott Vascular.

Event Text Entries

[185729114] The device is expected to be returned for evaluation. It has not yet been received. A follow up report will be submitted with all additional relevant information.
Patient Sequence No: 1, Text Type: N, H10


[185729115] It was reported that the procedure was performed to treat a heavily calcified lesion in the anterior tibial artery. A ht command 18 guide wire failed to cross due to resistance with the anatomy. During removal of the guide wire resistance was felt; however, it was unknown with what. Once the guide wire was almost completely out of the 4f sheath, the tip separated. The separated tip portion was at the exit of the sheath and was simply withdrawn. The procedure was successfully completed with a new ht command 18 guide wire. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-03082
MDR Report Key9898009
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-10-31
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 2024168
Manufacturer Street26531 YNEZ ROAD
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal Code925914628
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHT COMMAND 18
Generic NameGUIDE WIRE
Product CodeDQX
Date Received2020-03-30
Model Number1013731
Catalog Number1013731
Lot Number9103161
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report