MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-30 for XIENCE XPEDITION 1070275-48 manufactured by Abbott Vascular.
[188672760]
The device was not returned for evaluation. The reported patient effect of dissection is listed in the xience xpedition everolimus eluting coronary stent system instructions for use as a known patient effect of coronary stenting procedures. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. The reported patient effect of dissection is listed in the xience xpedition everolimus eluting coronary stent system instructions for use as a known patient effect of coronary stenting procedures. A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined. There is no indication of a product quality issue with respect to manufacture, design or labeling.
Patient Sequence No: 1, Text Type: N, H10
[188672761]
It was reported that the procedure was performed to treat a moderately calcified, moderately tortuous de novo left circumflex lesion that was 90% stenosed. A 2. 75x48mm xience xpedition stent was deployed and a distal edge dissection occurred. A new unspecified stent was deployed to cover the dissection and successfully complete the procedure. There was no adverse patient sequela reported. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2024168-2020-03085 |
MDR Report Key | 9898055 |
Report Source | COMPANY REPRESENTATIVE,DISTRI |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-02-18 |
Date Mfgr Received | 2020-03-09 |
Device Manufacturer Date | 2019-05-30 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. LINDSEY BELL |
Manufacturer Street | 26531 YNEZ RD. |
Manufacturer City | TEMECULA CA 925914628 |
Manufacturer Country | US |
Manufacturer Postal | 925914628 |
Manufacturer Phone | 9519143996 |
Manufacturer G1 | ABBOTT VASCULAR, REG # 3005718570 (P099) |
Manufacturer Street | CASHEL ROAD |
Manufacturer City | CLONMEL TIPPERARY |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | XIENCE XPEDITION |
Generic Name | DRUG ELUTING CORONARY STENT DELIVERY SYSTEM |
Product Code | NIQ |
Date Received | 2020-03-30 |
Catalog Number | 1070275-48 |
Lot Number | 9052342 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT VASCULAR |
Manufacturer Address | 26531 YNEZ RD. TEMECULA CA 925914628 US 925914628 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-30 |