MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-03-30 for R SERIES DEFIBRILLATOR manufactured by Zoll Medical Corporation.
[185540265]
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
Patient Sequence No: 1, Text Type: N, H10
[185540266]
Complainant alleged that during a routine shift check by a clinician, the device displayed a system fault 37 message. Complainant indicated that there was no patient involvement in the reported malfunction.
Patient Sequence No: 1, Text Type: D, B5
[187297801]
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to a faulty integrated circuit on the digital board. The device was recertified and returned to the customer. Analysis for reports of this type has not identified an increase in trend.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1220908-2020-00795 |
MDR Report Key | 9898120 |
Date Received | 2020-03-30 |
Date of Report | 2020-03-09 |
Date Mfgr Received | 2020-03-09 |
Device Manufacturer Date | 2014-06-01 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 269 MILL ROAD |
Manufacturer City | CHELMSFORD MA 01824 |
Manufacturer Country | US |
Manufacturer Postal | 01824 |
Manufacturer Phone | 9784219552 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | R SERIES DEFIBRILLATOR |
Generic Name | DEFIBRILLATOR/PACEMAKER |
Product Code | MKJ |
Date Received | 2020-03-30 |
Returned To Mfg | 2020-03-12 |
Model Number | R SERIES |
Catalog Number | R SERIES |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ZOLL MEDICAL CORPORATION |
Manufacturer Address | 269 MILL ROAD CHELMSFORD MA 01824 US 01824 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |