[186784409] This product is not marketed in the us. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186784410] The reporter indicated that a 12. 6mm, vticmo12. 6, -10. 0/1. 0/075 (sphere/cylinder/axis), implantable collamer lens, was implanted into the patients left eye (os) on (b)(6) 2019. On (b)(6) 2019 the lens was exchanged for a shorter length lens due to excessive vault. This exchange resolved the problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2023826-2020-00646
MDR Report Key9898226
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2020-03-30
Date of Report2020-03-03
Date of Event2019-11-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2019-08-14
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CYNTHIA THAI
Manufacturer Street1911 WALKER AVENUE
Manufacturer CityMONROVIA, CA
Manufacturer CountryUS
Manufacturer G1STAAR SURGICAL COMPANY
Manufacturer Street1911 WALKER AVENUE
Manufacturer CityMONROVIA, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMPLANTABLE COLLAMER LENS (ICL)
Generic NamePHAKIC TORIC INTRAOCULAR LENS
Product CodeQCB
Date Received2020-03-30
Model NumberVTICMO12.6
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTAAR SURGICAL COMPANY
Manufacturer Address1911 WALKER AVENUE MONROVIA, CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-30