[185600978] (b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185600979] According to the reporter, the patient died on (b)(6) 2020 and the device had not alarmed and presented the error message "no data for reported parameter". In a follow up phone conversation (b)(6) 2020 with the hospital technician it was clarified that the nurse manager of the orthopedic unit/floor was attempting to gather patient data but was unable to. The tech support team was able to find alarm reports for the dates of (b)(6) 2020 to (b)(6) 2020, but there was no information after that. The vital sync system showed that the patient was "discharged" by a nurse on (b)(6) 2020. They do not believe that the vital sync malfunctioned in any way since the patient was discharged from the vital sync system days prior to the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2936999-2020-00234
MDR Report Key9900478
Report SourceUSER FACILITY
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-05
Date Mfgr Received2020-03-10
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAVI LUGER
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone3035306582
Manufacturer G1AVNET INC
Manufacturer Street2740 AMERICAN BLVD W, STE 150
Manufacturer CityBLOOMINGTON MN 55431
Manufacturer CountryUS
Manufacturer Postal Code55431
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITAL SYNC REMOTE MONITORING SYSTEM
Generic NameTRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY
Product CodeDRG
Date Received2020-03-30
Model Number10116080
Catalog Number10116080
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerAVNET INC
Manufacturer Address2740 AMERICAN BLVD W, STE 150 BLOOMINGTON MN 55431 US 55431


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-03-30

© 2026 FDA.report