MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-30 for VIVA? QUAD XT CRT-D DTBA1Q1 manufactured by Medtronic Puerto Rico Operations Co..
[186764179]
Concomitant medical products: 7120 lead; implanted (b)(6) 2009. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186764180]
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited premature battery depletion. The device was explanted and replaced. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
[187144840]
Product event summary: the device was returned and analyzed. Returned product analysis was performed and no anomalies were found. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3004209178-2020-06405 |
MDR Report Key | 9900809 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-03-12 |
Date Mfgr Received | 2020-03-12 |
Device Manufacturer Date | 2016-09-16 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Street | ROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK |
Manufacturer City | JUNCOS PR 00777 |
Manufacturer Country | US |
Manufacturer Postal Code | 00777 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VIVA? QUAD XT CRT-D |
Generic Name | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO |
Product Code | NIK |
Date Received | 2020-03-30 |
Returned To Mfg | 2020-03-17 |
Model Number | DTBA1Q1 |
Catalog Number | DTBA1Q1 |
Device Expiration Date | 2018-03-14 |
Operator | LAY USER/PATIENT |
Device Availability | R |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Address | ROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization; 2. Required No Informationntervention | 2020-03-30 |