MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-30 for QUADRA ASSURA CRT-D QUAD RF HV CD3365-40Q manufactured by St. Jude Medical, Inc.(crm-sylmar).
[185699296]
The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10
[185699297]
It was reported that the implantable cardioverter defibrillator exhibited fine ventricular fibrillation with an isolated episode of undersensing. The patient was reported to be in stable condition.
Patient Sequence No: 1, Text Type: D, B5
[186482106]
New information received from the facility stated that the isolated undersensing episode occurred when the patient presented to the emergency room during ventricular fibrillation episode. The patient was externally paced causing the undersensing event observed. The device was reprogrammed.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2017865-2020-04409 |
MDR Report Key | 9901099 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-30 |
Date of Report | 2020-04-02 |
Date of Event | 2020-03-16 |
Date Mfgr Received | 2020-03-31 |
Device Manufacturer Date | 2013-12-04 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | ELIZABETH BOLTZ |
Manufacturer Street | 15900 VALLEY VIEW COURT |
Manufacturer City | SYLMAR CA 91342 |
Manufacturer Country | US |
Manufacturer Postal | 91342 |
Manufacturer G1 | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) |
Manufacturer Street | 15900 VALLEY VIEW COURT |
Manufacturer City | SYLMAR CA 91342 |
Manufacturer Country | US |
Manufacturer Postal Code | 91342 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | QUADRA ASSURA CRT-D QUAD RF HV |
Generic Name | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |
Product Code | NIK |
Date Received | 2020-03-30 |
Model Number | CD3365-40Q |
Catalog Number | CD3365-40Q |
Lot Number | 4373320 |
Device Expiration Date | 2015-11-30 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) |
Manufacturer Address | 15900 VALLEY VIEW COURT SYLMAR CA 91342 US 91342 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |