HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM, JP 104911JPN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2020-03-30 for HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM, JP 104911JPN manufactured by Thoratec Corporation.

Event Text Entries

[186295139] No further information was provided. A supplemental report will be submitted when the manufacturer? S investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[186295140] It was reported that the patient heard a red heart alarm. When the alarm occurred the patient had syncope. Upon log file analysis, pump stops and low speed operation were found. The patient was admitted and connected to battery power. The patient underwent a pump exchange on (b)(6) 2020 due to the driveline damage.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916596-2020-01623
MDR Report Key9901273
Report SourceDISTRIBUTOR,FOREIGN
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-18
Date Mfgr Received2020-03-18
Device Manufacturer Date2017-09-19
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BOB FRYC
Manufacturer Street6035 STONERIDGE DRIVE
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal94588
Manufacturer Phone7818528204
Manufacturer G1THORATEC CORPORATION
Manufacturer Street6035 STONERIDGE DRIVE
Manufacturer CityPLEASANTON CA 95488
Manufacturer CountryUS
Manufacturer Postal Code95488
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM, JP
Generic NameVENTRICULAR (ASSISST) BYPASS
Product CodeDSQ
Date Received2020-03-30
Model Number104911JPN
Lot Number6161219
Device Expiration Date2019-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC CORPORATION
Manufacturer Address6035 STONERIDGE DRIVE PLEASANTON CA 95488 US 95488


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-30

© 2026 FDA.report