MITRACLIP? SYSTEM CLIP DELIVERY SYSTEM CDS0602-NTR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-30 for MITRACLIP? SYSTEM CLIP DELIVERY SYSTEM CDS0602-NTR manufactured by Abbott Vascular.

Event Text Entries

[185726719] The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot. Additionally, a review of the complaint history identified no other similar incidents reported from this lot. All available information was investigated, a definitive cause for the reported unintended movement (clip open? During establishing final arm angle/efaa) could not be determined. There is no indication of a product quality issue with respect to manufacture, design or labeling. The other mitraclip referenced is filed under a separate medwatch report number.
Patient Sequence No: 1, Text Type: N, H10


[185726720] This is filed to report the clip failed the final arm angle test. It was reported that this was a mitraclip procedure performed to treat degenerative mitral regurgitation (mr), with an mr grade of 3. The transseptal puncture was high, 5. 5cm. The clip delivery system (cds) (91019u222) was advanced to the mitral valve. After grasping the leaflets, the clip failed the final arm angle test twice. The clip was not implanted and was removed. Another clip (91106u183) was advanced, but the clip slipped too medial and got caught in chordae. Troubleshooting was performed and the clip was freed; however, a chordae rupture occurred. The clip was implanted, and mr reduced. To further reduce mr another clip was implanted, reducing mr to 1. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-03113
MDR Report Key9901290
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-10-21
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005070406
Manufacturer Street3885 BOHANNON DRIVE
Manufacturer CityMENLO PARK CA 94025
Manufacturer CountryUS
Manufacturer Postal Code94025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMITRACLIP? SYSTEM CLIP DELIVERY SYSTEM
Generic NameVALVE REPAIR
Product CodeNKM
Date Received2020-03-30
Catalog NumberCDS0602-NTR
Lot Number91019U222
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.