PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM 12673-03

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-30 for PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM 12673-03 manufactured by Abbott Vascular.

Event Text Entries

[185730087] The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy, suture pulled until it breaks or tensioning angle is not being coaxial due to circumstances of the procedure. There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
Patient Sequence No: 1, Text Type: N, H10


[185730088] It was reported that an arteriotomy closure of the left common femoral artery was attempted using a proglide device with a 6f sheath after an interventional procedure in the popliteal and tibia arteries. Reportedly, when harvesting the suture outside the patient it was noticed that the suture was frayed and broke. The remaining portion of the suture was still able to be used to achieve hemostasis. There was no adverse patient effect and no reported clinically significant delay in the procedure or therapy. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-03122
MDR Report Key9901454
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-09
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-12-09
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005718570 (P099)
Manufacturer StreetCASHEL ROAD
Manufacturer CityCLONMEL TIPPERARY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Generic NameSUTURE MEDIATED CLOSURE
Product CodeMGB
Date Received2020-03-30
Model Number12673-03
Catalog Number12673-03
Lot Number9120941
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.