8CH INFINITY DBS FLEX EXTN KIT, 60CM, B 6372

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-30 for 8CH INFINITY DBS FLEX EXTN KIT, 60CM, B 6372 manufactured by St. Jude Medical - Neuromodulation.

Event Text Entries

[185806337] Microscopic inspection of the returned extension identified 2 fractures sites where multiple wires were broken. Continuity and impedance tests confirmed all the channels were electrically open. This would have contributed to the patient? S loss of therapy. As the high impedance was observed prior to the revision, the fractures are consistent with the lead being subjected to an overstress condition while in vivo. The source of the stress could not be conclusively determined. Date of event is estimated.
Patient Sequence No: 1, Text Type: N, H10


[185806339] It was reported the patient experienced a return of symptoms. Patient indicated when he increased stimulation strength, the amplitude would automatically reduce. Diagnostics indicated 7 out of 8 contacts showed high impedance readings. Reprogramming was unable to provide resolution. Surgical intervention was undertaken and the extension was explanted and replaced. No complications were noted during the procedure. Postoperative, the patient is reportedly receiving effective therapy.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1627487-2020-03459
MDR Report Key9901528
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2019-06-06
Date Mfgr Received2019-12-16
Device Manufacturer Date2018-07-18
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANDREA DEITZ
Manufacturer Street6901 PRESTON ROAD
Manufacturer CityPLANO TX 75024
Manufacturer CountryUS
Manufacturer Postal75024
Manufacturer Phone9723098000
Manufacturer G1ST. JUDE MEDICAL - NEUROMODULATION
Manufacturer Street6901 PRESTON RD
Manufacturer CityPLANO TX 75024
Manufacturer CountryUS
Manufacturer Postal Code75024
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name8CH INFINITY DBS FLEX EXTN KIT, 60CM, B
Generic NameDBS EXTENSION
Product CodeMHY
Date Received2020-03-30
Returned To Mfg2020-01-07
Model Number6372
Catalog Number6372
Lot Number6546698
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL - NEUROMODULATION
Manufacturer Address6901 PRESTON RD PLANO TX 75024 US 75024


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.