MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-30 for ACCOLADE DR L301 manufactured by Boston Scientific Corporation.
[185775891]
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
Patient Sequence No: 1, Text Type: N, H10
[185775892]
It was reported that the battery life of this pacemaker went from eight years remaining to five and a half years remaining in a span of seven months. The engineering analysis revealed that the device was at high atrial sensing and has had several atrial tachy response (atr) episodes. The device also had signal artifact monitoring (sam) patched installed and enabled. The boston scientific technical services (ts) recommended reprogramming. The device remains in service. No adverse patient effects were reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2124215-2020-04238 |
MDR Report Key | 9901572 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2020-03-30 |
Date of Report | 2020-03-30 |
Date of Event | 2020-01-22 |
Date Mfgr Received | 2020-01-22 |
Device Manufacturer Date | 2018-05-28 |
Date Added to Maude | 2020-03-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | TIMOTHY DEGROOT |
Manufacturer Street | 4100 HAMLINE AVENUE NORTH |
Manufacturer City | SAINT PAUL MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 6515826168 |
Manufacturer G1 | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Street | CASHEL ROAD |
Manufacturer City | CLONMEL |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ACCOLADE DR |
Generic Name | IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) |
Product Code | LWP |
Date Received | 2020-03-30 |
Model Number | L301 |
Catalog Number | L301 |
Lot Number | 769769 |
Operator | LAY USER/PATIENT |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Address | 4100 HAMLINE AVENUE NORTH SAINT PAUL MN 55112 US 55112 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-30 |