INOGEN CRT-D G140

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-30 for INOGEN CRT-D G140 manufactured by Boston Scientific Corporation.

Event Text Entries

[185747154] It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited high thresholds and high out-of-range (oor) pacing impedances on the left ventricular (lv) lead, measuring greater than 2000 ohms. It was noted the lv lead was capturing and sensing appropriately. The lead configuration was reprogrammed. This crt-d and lv lead remain in service. No adverse patient effects were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2124215-2020-01609
MDR Report Key9901593
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-01-15
Date Mfgr Received2020-02-04
Device Manufacturer Date2015-04-23
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY DEGROOT
Manufacturer Street4100 HAMLINE AVENUE NORTH
Manufacturer CitySAINT PAUL MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone6515826168
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetCASHEL ROAD
Manufacturer CityCLONMEL
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINOGEN CRT-D
Generic NameDEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Product CodeNIK
Date Received2020-03-30
Model NumberG140
Catalog NumberG140
Lot Number478469
Device Expiration Date2017-04-23
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address4100 HAMLINE AVENUE NORTH SAINT PAUL MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report