RIATA ST PASSIVE FIXATION 7040/65

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-30 for RIATA ST PASSIVE FIXATION 7040/65 manufactured by St. Jude Medical, Inc.(crm-sunnyvale).

Event Text Entries

[185715195] The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185715196] It was reported that the right ventricular lead exhibited non sustained right ventricular oversensing and noise. Lead revision was discussed. The patient was stable.
Patient Sequence No: 1, Text Type: D, B5


[188578664] New information received notes that the lead was capped and replaced on (b)(6) 2020. The patient was stable before, during and after the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2938836-2020-01957
MDR Report Key9901630
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-30
Date of Report2020-03-30
Date of Event2020-03-12
Date Mfgr Received2020-03-12
Device Manufacturer Date2006-06-06
Date Added to Maude2020-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELIZABETH BOLTZ
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer G1ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Street645 ALMANOR AVENUE
Manufacturer CitySUNNYVALE CA 94085
Manufacturer CountryUS
Manufacturer Postal Code94085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRIATA ST PASSIVE FIXATION
Generic NameDEFIBRILLATION LEAD
Product CodeNVY
Date Received2020-03-30
Model Number7040/65
Lot Number0002113713
Device Expiration Date2009-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Address645 ALMANOR AVENUE SUNNYVALE CA 94085 US 94085


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.