EDGE E1455-4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for EDGE E1455-4 manufactured by Covidien Mfg Dc Boulder.

Event Text Entries

[185656161] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185656162] According to the reporter, during bilateral needle localisation biopsy and central node biopsy, the patient had a 6cm away from the surgical site a full thickness 3rd degree burn requiring to be excised. The surgeon excised the burn site and sutured it. The surgeon described that there was a hole in electrode body due to scrub nurse changing electrode tip with artery. A photo submitted showing the long electrode blue insulation has small hole in the middle of its length.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1717344-2020-00359
MDR Report Key9902237
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-12
Date Mfgr Received2020-03-12
Device Manufacturer Date2019-11-21
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA HERNANDEZ
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone2034925563
Manufacturer G1COVIDIEN MFG DC BOULDER
Manufacturer Street5920 LONGBOW DR
Manufacturer CityBOULDER CO 803013299
Manufacturer CountryUS
Manufacturer Postal Code803013299
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEDGE
Generic NameELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Product CodeGEI
Date Received2020-03-31
Model NumberE1455-4
Catalog NumberE1455-4
Lot Number93160126X
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN MFG DC BOULDER
Manufacturer Address5920 LONGBOW DR BOULDER CO 803013299 US 803013299


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.